IUV-25CP
Global Site> Irrdiation System> IUV-25CP Spectral Irrdiation System For Photoresist 

IUV-25CP  Spectral Irrdiation System For Photoresist

This is the system that multiple monochromatic lights in a wavelength band (approx.240 nm) is irrdiated to photosensitive material such as photoresist agest for arbitraty time. Two types of light source are equipped as standard. One is the Deep UV lamp which can irrdiate strong line spectrum light and another is Xe lamp which can irradiate continuous spectrum light.

The sample can be moved stepwise on the automatic stage, and it can be irradiated while changing irradiation time.

 Catalog(PDF)

 

Features

 

  • Testing the spectral sensitvity (wavelength characteristics ) of photoresist agents can be performed.

 

  • The travel distance and the number of steps can be set to 5mm×13 stpes at maximum(2mm is required between irradiation areas )

 

  • The irradiation time and position can be set and controlled with the notebook PC and software.

 

  • The Deep UV lamp can be changed to an ultrahigh pressure mercury lamp as an option

 

 

Image of Irrdiation

 

Holder ( 2 types )

 

 

 

Software

 

  • Start ( condition ) : Enter the first seconds of irradiation time ( 0.01 to 1200 )
  • End : Enter the last seconds of irrdiation time ( 0.01 to 1200 )
  • Number of divisions :Enter the number of division between start and end (2 to 13 )
  • Division method : Select from Linear or Log
  • Calculation : Set the irradiation time (seconds) under specifled conditions and display it in the list
  • Close : Close the irradiation conditions screen and returns to the main menu
  • Manual : Set shutter state and stage position arbitarily
  • Start (whole ) : Start Irrdiation under the set conditions

 

 

Specifications

 

 Irrdiation Wavelength range  Wavelength band irrdiated at once : about 240nm 
 Irrdiation area  Long distance travel type. Height direction : 89mm (5×13 steo.space btween irradiation:2mm)
   WL direction : approx.80mm (approx.240nm)
 Irrdiation time  0.01~1200sec
 Min. setting time  0.01sec
 Irrdiation step  1~13 steps
 Direction fo divistion of irradiation  Select either Linear or log
 Light source

 DeepUV lamp 500W・Xe lamp 500W

 Lamp swithcing  Manural swithing with the mirror
 Sample holder  □100mm、□127mm for glass plate ( exchange type )

 

 

Standard Configurations

 

●Deep UV lamp 500W
●Deep UV lamp holder(air cooling type)
●Deep UV lamp power supply
●Xe lamp 500W
●Xe lamp holder(air cooling type)
●Xe lamp power supply
●Light condensing optical system(manual swithing of 2 light paths)
●Auto shutter mechanism
●M25-CP Monochromator
●Grating 1200lines/300nmBlaze(MgF2 coating)
●Sample chamber(auto stage)
●Sample holder for glass sample,100×100mm
●Sample holder for glass sample,127×127mm
●Referebce detector (silicon photon diode with certificate)
●Ammeter for signal of reference detector (Electro meter)
●Stage controller
●Shutter controller
●Note PC
●Dedicated software
●Instruction Manual


 

Outer Dimensions

 

●Main unit  :approx. W1000×D650×H470mm
        *Excluding lamp power suppley, controller, electro-meter

 

 

Related Product

 
NIJI-2 Variable Wavelength Light Source

 
  • Irradiation in the wide range from ultraviolet,visble,to Infrared light can becoverd
  • Stand-alone and compact desigh require only a modest requirement of bench space for installation 

 

MM-3  Multi Waveleght Irradiation System

 
  • It is possible to test deterioration dependence on wavelenth for various samples
  • The Model MM-3 can irradiate the sample in arbitrary wavelenght range in the wavelength rage of 220~770nm

 

BIR-50  Solar Cell Light Resistance Testing System

 
  • Long accurate light degradation test with the solar simulator
  • I-V measurement under constant temperature atmosphere(20~90℃)